USP 35 Carisoprodol tablets.
HPLC analysis for assay.
asked to calculate by using heights.
In second supplent, it is revised to peak area.
Using peak area is a good practice for assays.
Using peak heights is a good choice when two adjascent peaks are not seperated completely.
Regards,
Bujji Kanchi.
8008889185.
Friday, 21 December 2012
How to understand time required to Incubate for photostability testing
How much duration of time to hold the samples in a Photostability chamber.
ICH Defines, Expose only one batch is enough to 1.2 million lux hours(This is unlike the frequency of stability testing, eg: months 3,6,9,12.. etc)
How to convert this(1.2 million lux hr.) into time or days.
Explanation :
Lux is the unit of illumunation( )
[Lux is both singular and plural like the word "water"]
Let us assume that the intensity at particular location in the chamber for test sample is 4430 lux.
For the test sample to get exposed to 1.2 million lux hr, it has to be kept in photostability chamber for the following number of days
Y = 1.2 x 106 lux hr / 4430 lux = 270.88 hr = 271 hr
Converting hours into number of days = 271 hr/24 hr = 11.29 days ~ 11.5 days.
ICH Q1B Requirement :
For confirmatory studies, samples should be exposed to light providing an overall illumination of not less than 1.2 million lux hours and an integrated near ultraviolet energy of not less than 200 watt hours/square meter to allow direct comparisons to be made between the drug substance and drug product.
FYI,
Regards,
Bujji Kanchi.
Subscribe to:
Posts (Atom)
Stability of drug substances
Stability studies : Define stability : To study on how the product quality changes with time under the influence of temperature, humi...
-
Water determination Reference USP General Chapter<921> Simple Reaction : water against KF Reagent(Karlfischer reagent) ROH...
-
Ho w much duration of time to hold the samples in a Photostability chamber. ICH Defines, Expose only one batch is enough to 1.2 mill...
-
Dirty equipment Hold time studies : Now a days the auditors are asking about the Dirty equipment Hold time study report in cleaning va...