Thursday, 29 November 2018

Air handling unit (AHU) Qualilfication and Clean room classification


Air handling unit (AHU) Qualilfication  and clean room Classification : 



 Ensure the AHU is in switch ON.
 Switch ON the anemometer and allow it to some time for stabilization.
  Ensure that anemometer measuring unit is in FPM mode.
 To change the measuring unit mode, press the UNIT button till display shows FPM.
Take the readings at five locations at each filter as per below mentioned diagram.
           

V1                                    V2

                     V5

V3                                    V4
 



  •  Position the probe pointing towards airflow. Locate the probe approximately 1/6 diagonal distance and measure velocities from each corner.For clean room measurement shall be taken at a distance of 150 mm from the filter face. Take the readings after stabilization or take the average reading of oscillating value shown in the display.

Procedure to Calculate the Air changes
  Calculate the supply grill area A= Length (in ft) x Width (in ft).
Calculate the supply grill Average velocity V= (V1+V2+V3+V4+V5 ) /5.( in ft/min) .
Calculate the air flow delivered by each supply grill in CFM   = A x V (in CFM).
Find the total quantity of supply air delivered in to the room in CFM by adding all             supply grills air flow CFM.
Calculate the air changes by below given formula.
                    Air changes per hour = Total supply air flow rate (CFM) in the clean room x 60
                                                                        Volume of the clean room (cu. ft)          




Example :
Air velocity =  100 feet per min
Air Flow =  Velocity Ft/ min x grill area(Ft2)
                              =  100 x 4 =400 Cubic feet per min(CFM)

If room has two supply grills, sum the two air flows = 400+ 400= 800 CFM

Air changes = Air flow x 60
Room volume(Ft3 )

=800 Ft3x 60 min
  1000 Ft3

= 48 Air changes per hour (Limit : Not less than 25 air changes per hour for ISO 8)




There are 5 tests during AHU Qualification.

S.NO
                   Test
       Frequency
Instruments used
Acceptance criteria
1
Air Changes
For every six months
(on or before 7 days)
Anemometer for Velocity of air
Not less than 25 air changes per hour
2
Particle count “At Rest Condition”
Particle counter
Not more than 3520000 for ISO 8
There is an equation to estimate particle limit.
3
HEPA Filter Integrity Test
(Leak Test)
Aerosol Photometer and Aerosol generator using compressed air
No leak should be observed
Not more than 0.01% leak
4
Recovery test
For every one year
(on or before 15 days)
Particle counter
Recovered within 15 minutes, after polluting the room with aerosols
5
Air flow pattern
Dry ice in hot water to generate SMOKE
Uni-directional and positive and negative pressure should be indicative.



Regards,
Bujji Reddy.
9676749937














     

10 comments:

  1. https://www.blogger.com/comment.g?blogID=3915901706881318158&postID=3634607346034562518&page=1&token=1587663474403&isPopup=trueCleanroom machinery Relocation

    ReplyDelete
  2. Very interesting content about software testing services and it's advantages. Testing playing a major role software development. Here you can check more about software testing services.

    ReplyDelete
  3. Great post! I really enjoyed reading your thoughts on this topic.Thanks for sharing!!
    Ayurvedic PCD Pharma Franchise in Rajasthan

    ReplyDelete
  4. Really Nice Information It's Very Helpful Thanks for sharing such an informative post! India's best Ayurvedic PCD franchise companies

    ReplyDelete
  5. Really Nice Information It's Very Helpful Thanks for sharing such an informative post! Top Ophthalmic PCD Franchise in India

    ReplyDelete

Stability of drug substances

Stability studies : Define stability  : To study on how the product quality changes with time under the influence of temperature, humi...